Consort

Consort develops a wide range of high technology products from simple and robust field-meters up to sophisticated precision laboratory equipment. There are our well-known meters and controllers for pH, mV, ion analysis, rH2, temperature, conductivity, resistivity, salinity, TDS and dissolved oxygen. For the molecular biology field, we offer a wide range of electrophoresis equipment such as high voltage power supplies, horizontal and vertical apparatus, DNA sequencers, blotting units, UV-tables and photo-documentation system.

consort_prolyse

Dataloggers

A Data Logger (also Datalogger or Data Recorder) is an electronic device that records data over time or in relation to location either with a built in instrument or sensor or via external instruments and sensors. Increasingly, but not entirely, they are based on a digital processor (or computer). They generally are small, battery powered, portable, and equipped with a microprocessor, internal memory for data storage, and sensors


climate_zone_enclosure_prolyse

D230

multi-channel data-logger for pH, mV, ion, conductivity, TDS, dissolved oxygen and 4…28 pH/mV/Ion channels

4…28 conductivity channels

4…28 oxygen channels

4…28 temperature channels

Built in a standard 19 inch rack + cabinet the central unit accepts up to 7 modules. Different modules can be mixed in one unit.
There are two modules available:
D291: 4 pH/mV/Ion/dissolved oxygen + 4 temperature channels,
D292: 4 conductivity/TDS + 4 temperature channels.
Up to 16 units can be combined with each other to increase the number of channels up to 448.
All parameters including set-up, calibration and display are controlled through any computer via a single RS232 connection.
All necessary GLP-procedures are on board. The software package supplied with the D230 is specially designed to control, collect and store data. Runs under Windows 2000 or higher. All channels are processed at the same time, each in its own window. The software automatically detects the maximum number and type of available channels.
Data is collected on-line with a programmable interval (4 s … 24 h).
Documented printouts will show file name, date and time, name of the operator, name of the company, name of the division, optional notes by the operator.
Password protection prevents any unauthorised access to the instrument.

D291
pH: 0…14 pH
mV: ± 2000 mV
Ion: 0.01 ng/l…99.9 g/l
Dissolved oxygen: 0…60 mg/l, 0…600%
Temperature: -30…+130°C
Resolution: 0.001 pH, 0.1 mV, 0.01 mg/l, 0.1°C

D292
Conductivity: 0…2000 mS/cm
TDS: 0…100 g/l
Resolution: 0.01µS/cm, 0.01 mg/l, 0.1°


climate_zone_enclosure_prolyse

C3060 – C3061

multi-channel data-logger for pH, mV, ion, conductivity, resistivity, salinity, TDS, dissolved oxygen and temperature

8 independent channels for all measurements!
(conductivity and temperature: only 2 channels)

A large bright LCD screen with white backlight enables to view all channels simultaneously.
Multi-point (1…5) calibration of pH electrodes with any of eleven pre-programmed and five user specified buffer tables.
Multi-point (1…3) calibration of conductivity electrodes for increased linearity with any of three pre-programmed or user specified standard tables.
Multi-point (2…5) calibration + blank correction of ion-selective electrodes.
The display guides you at any moment in the language of your choice (English, Dutch, French, German).
All necessary GLP-procedures are on board.
Password protection prevents any unauthorised access to the instrument.
Three year warranty.
Storage memory for up to 12000 values + date and time.
Two-way communication with any computer using USB (C3060).
Two-way communication with any computer using Ethernet (C3061). NEW !

Specifications
pH: -2…+16 pH
mV: ± 2000 mV
Ion: 10-11…10+1, any unit
Conductivity: 0…2000 mS/cm
Resistivity: 0…200 Mohm.cm
Salinity: 0…70 ppt
TDS: 0…100 g/l
Temperature: -5…+105°C
Resolution: 0.001 pH, 0.1 mV, 0.001 µS/cm, 1 ohm.cm, 0.01 mg/l, 0.1 ppt, 0.1°C


climate_zone_enclosure_prolyse

DIS-1

data acquisition software for meters or controllers equipped with a USB, Ethernet or RS232 interface

This software package is specially designed to collect and store data from the following models:
C3010, C3011, C3020, C3021, C3030, C3031, C3040, C3041, C3050, C3051, C3060, C3061.
C931, C932, C933, K912.
T871, T872.
It can also be used with the following previous models or versions:
C861, C862, C863, C864, C868.
C831, C832, C833, C834, C835, C838, C731, C732, C733.
P901, P902, P903, K911, Z921.
R305, R315, R325, R335, R735.
T851, T852.
Runs under Windows 2000 or higher.
Allows to log, view and print the incoming values in real time or to retrieve data-logged values from the memory of the instrument.
All channels of multi-channel instruments can be processed simultaneously.
Values are shown in separate windows, tables or graphs.
A graph window also allows to set alarm limits and to view several statistic values.
Guides you at any moment in the language of your choice (English, Dutch, French).


climate_zone_enclosure_prolyse

DIS-2

data acquisition software for controllers equipped with a RS485 interface

Special software to transfer data from the following models to a personal computer:
R362, R363
It can also be used with the following previous models or versions:
R305, R315, R335.
Runs under Windows 2000 or higher.
Allows to log, view and print the incoming values in real time or to retrieve data-logged values from the memory of the instrument.
Up to 31 controllers can be processed simultaneously.
Values are shown in separate windows, tables or graphs.
A graph window also allows to set alarm limits and to view several statistic values.
Guides you at any moment in the language of your choice (English, Dutch, French).